Precise Monitoring · Smart Equipment Protection · Empowering Operations and Maintenance
The YGF-8 Oil Spectrometer is a high-precision analytical instrument based on the principle of Atomic Emission Spectroscopy (AES) and utilizing Graphite Rotating Disc Electrode (RDE) technology. It can quickly and accurately detect the content of metal elements in various oils such as lubricating oil, fuel oil, and hydraulic oil, covering wear metals, contaminants, and additive-related elements. It provides scientific, reliable data support for equipment condition monitoring, fault diagnosis, and maintenance decision-making.
The YGF-8 Oil Spectrometer acts as the "perspective lens" for equipment wear and the "gold standard" for oil monitoring. By utilizing advanced Atomic Emission Spectroscopy (AES) and Graphite Rotating Disc Electrode (RDE) technology, it delivers rapid, highly accurate detection of metal elements in lubricating oils, fuels, and hydraulic fluids.
With a detection limit below 1 ppm and a single test time of ≤35 seconds, it eliminates the need for complex sample digestion or pretreatment. The system is equipped with a proprietary spectral analysis expert system that supports data management, trend analysis, and fault warning, empowering predictive maintenance and minimizing downtime risks.
| Parameter | Value |
|---|---|
| Detection Principle | Atomic Emission Spectroscopy (AES), Graphite Rotating Disc Electrode |
| Optical System | High-performance CMOS, 4096-pixel resolution, 3-fiber input |
| Optical Temperature Control | Constant temperature 38℃ ± 0.1℃ |
| Spectral Line Range | Wide range: 200 ~ 850 nm |
| Detection Elements | Standard 24+ elements (Al,Cr,Cu,Fe,Pb,Mn,Mo,Ni,Ag,Sn,B,Na,Si,K,Li,V,Ba,Ca,Mg,Ti,Zn,As,Bi,Ce,Co,In,Sb,Se,Sr,W,Zr, etc.), supports flexible expansion |
| Detection Range | 0 ~ 1000 ppm (19 elements), 0 ~ 6000 ppm (5 elements) |
| Detection Limit | ≤ 1 ppm |
| Single Detection Time | ≤ 35 seconds |
| Sampling Volume | Approx. 2 mL |
| Parameter | Value |
|---|---|
| Excitation Power Supply | High-energy arc excitation, ignition voltage 14000V, digital discharge parameter setting |
| Hardware Configuration | Built-in industrial control computer, graphite electrode grinding machine; integrated design |
| Dimensions | L 700 mm × W 350 mm × H 550 mm |
| Weight | Approx. 46 kg (Net weight) |
| Power Requirements | 110V or 220V or adjustable, 50/60Hz AC, built-in voltage stabilizer (no dedicated grounding required) |
| Working Environment | Temperature: -40℃ ~ 50℃, adapted for stable operation in multiple scenarios |
| Software Functions | User-friendly interface, one-click detection, data management, trend analysis, fault warning, data export & auto storage |
| Execution Standards | ASTM D6595, ASTM D6728, NB/SH/T 0865, SN/T 1652, HB 20094.1, DL/T 1550 |
| # | Item | Qty. |
|---|---|---|
| 1 | YGF-8 Main Unit (with built-in industrial control computer) | 1 set |
| 2 | Graphite Electrode Grinding Machine (integrated) | 1 set |
| 3 | High-performance CMOS Optical Module (4096-pixel) | 1 set |
| 4 | Power Supply with Built-in Voltage Stabilizer | 1 set |
| 5 | Spectral Analysis Expert System Software (Pre-installed) | 1 license |
| 6 | Initial Consumables Kit (Graphite electrodes & sample cups) | 1 kit |
| 7 | User Manual & Compliance Certificates | 1 set |